PAMAS 4132
The PAMAS 4132 is used as a particle counting system for Single & Multipass Filter test rig (1 or 2 boxes). The standard system fits in the test rigs of filter manufacturers. The Single Pass Filter Test requires one instrument in the test rig. For the Multipass Filter Test two instruments in the test rig for simultaneously upstream and downstream measurements in 2 x 32 size channels are used.
Available sensors:
HCB-LD-50/50
Size range:
1 - 400 μm with latex calibration
1 - 100 μm with calibration according to ISO 4402
4 - 70 μm(c) with calibration according to ISO 11171
Max. particle concentration: 24,000 p/ml at 25 ml/min
HCB-LD-25/25
Size range:
1 - 200 μm with latex calibration
1 - 100 μm with calibration according to ISO 4402
4 - 70 μm(c) with calibration according to ISO 11171
Max. particle concentration: 120,000 p/ml at 10 ml/min
The system is adaptable to any Multipass application and ß-ratio analysis. Please contact us for integration into your application.